Quantitative
Overview of Quantitative Hemoglobin Measurement
- Definition: Quantitative hemoglobin (HGB) measurement refers to laboratory techniques used to determine the concentration of hemoglobin in a blood sample
-
Clinical Significance
- Diagnosis of Anemia: Decreased HGB levels indicate anemia
- Diagnosis of Polycythemia: Increased HGB levels indicate polycythemia
- Monitoring Treatment: Used to monitor the response to treatment for anemia or polycythemia
- Assessing Blood Loss: Helps estimate the severity of blood loss
- Evaluating Overall Health: Provides information about the oxygen-carrying capacity of the blood
- Method: Spectrophotometry is the primary method used for quantitative hemoglobin measurement in automated hematology analyzers
- Reporting Units: Hemoglobin concentration is typically reported in grams per deciliter (g/dL) or grams per liter (g/L)
Principle of Spectrophotometric Hemoglobin Measurement
Spectrophotometry relies on the Beer-Lambert Law, which states that the absorbance of a solution is directly proportional to the concentration of the analyte and the path length of the light beam through the solution:
- A = εbc
- A = Absorbance
- ε = Molar absorptivity (a constant specific to the substance)
- b = Path length (the distance the light beam travels through the solution)
- c = Concentration
In the context of hemoglobin measurement:
- The intensity of the color produced by the converted hemoglobin is directly proportional to the hemoglobin concentration
- The spectrophotometer measures the absorbance of the solution, and the hemoglobin concentration is calculated based on the Beer-Lambert Law
Detailed Steps of Spectrophotometric Hemoglobin Measurement
-
Sample Preparation
- Whole Blood Collection: Blood is collected in an EDTA (purple-top) tube to prevent clotting
- Lysis of Red Blood Cells (RBCs):
- A lysing reagent is added to the whole blood sample to rupture the RBC membranes and release hemoglobin into the solution
- The lysing reagent also clears the solution, reducing turbidity
- Different lysing reagents may be used, depending on the specific method used by the analyzer
- Conversion of Hemoglobin to a Stable Form:
- The released hemoglobin is converted to a stable, colored compound that can be accurately measured by spectrophotometry
- The most common method is the cyanmethemoglobin (hemiglobincyanide) method
-
Cyanmethemoglobin (HiCN) Method
- Potassium Ferricyanide:
- Converts hemoglobin to methemoglobin (Hi), in which the iron is in the ferric (Fe3+) state
- Potassium Cyanide:
- Reacts with methemoglobin to form cyanmethemoglobin (HiCN), a stable, colored compound
- The reaction is as follows: Hb + K3Fe(CN)6 → Hi + KCN → HiCN
- Absorbance Measurement:
- The solution is passed through a spectrophotometer, and the absorbance is measured at a specific wavelength (typically 540 nm)
- A blank (containing the lysing reagent but no blood) is used to zero the spectrophotometer
- The absorbance is directly proportional to the cyanmethemoglobin concentration, which in turn is directly proportional to the hemoglobin concentration in the original sample
- Calculation:
- The hemoglobin concentration is calculated using a calibration curve or a factor derived from the Beer-Lambert Law
- Potassium Ferricyanide:
-
Automated Hematology Analyzer Process
- Sample Aspiration: The automated analyzer aspirates a small amount of the prepared blood sample
- Mixing and Incubation: The analyzer mixes the sample with the lysing reagent and allows sufficient time for the reaction to occur
- Flow Cell: The solution is passed through a flow cell, where the light beam from the spectrophotometer passes through the solution
- Detection: A photodetector measures the amount of light that passes through the solution
- Calculation: The analyzer calculates the hemoglobin concentration based on the absorbance and calibration data
- Reporting: The hemoglobin concentration is displayed and reported on the analyzer’s printout or screen
Interfering Substances & Troubleshooting
Several substances can interfere with spectrophotometric hemoglobin measurement, leading to inaccurate results. It’s crucial to be aware of these interferences and take appropriate corrective actions
-
Turbidity: Lipemia (excessive lipids in the blood), high white blood cell counts (leukocytosis), or the presence of non-lysed red blood cells can cause turbidity, increasing the absorbance and falsely elevating the hemoglobin result
- Corrective Actions:
- Lipemia: Perform a saline replacement procedure or use a lipemia clearing agent
- High WBC Count: Dilute the sample and multiply the results by the dilution factor
- Non-Lysed RBCs: Ensure adequate mixing and incubation time with the lysing reagent
- Corrective Actions:
-
High White Blood Cell Count (Leukocytosis)
- Extreme leukocytosis can cause falsely elevated hemoglobin readings
- This is due to light scattering caused by the high concentration of cells
- Corrective Action:
- Dilute the sample with an isotonic diluent (e.g., saline) and repeat the measurement
- Multiply the result by the dilution factor
- A manual hemoglobin method can also be used to confirm the accuracy
-
Lipemia (High Lipid Levels)
- Turbidity from high lipid content can falsely elevate the hemoglobin reading
- Corrective Actions:
- Saline Replacement: Replace the lipemic plasma with an equal volume of isotonic saline after centrifugation
- Lipemia Clearing Agents: Use commercial reagents to clear the lipemia before analysis
- Ultracentrifugation: Remove the lipids by ultracentrifugation
-
Cold Agglutinins
- These antibodies can cause RBCs to clump together, leading to inaccurate cell counts and hemoglobin measurement
- Corrective Actions:
- Warm the sample to 37°C to dissociate the agglutinins
- Repeat the analysis promptly after warming
- Corrective Actions:
- These antibodies can cause RBCs to clump together, leading to inaccurate cell counts and hemoglobin measurement
Quality Control
-
Calibration
- Perform regular calibration of the spectrophotometer according to the manufacturer’s instructions
- Use certified reference materials (calibrators) with known hemoglobin concentrations
-
Control Materials
- Run control materials (low, normal, and high levels) at regular intervals (e.g., daily, with each batch of samples)
- Use controls that are appropriate for the types of samples being analyzed
-
Review Control Results
- Review control results and evaluate them using statistical methods (e.g., Levey-Jennings charts, Westgard rules)
- Take corrective action if control results are outside the acceptable range